Recall Z-1249-2009 ยท HRS Plate, Fixation, Bone

Synthes Spine

Synthes InSitu Bender Cutter Kit Narrow Sterile Catalog number 530.522S, used with Synthes Rapid Resorbable Fixation System for craniofacial skeleton fracture repair and reconstructive procedures.

DateJun 1, 2009 (initiated Dec 18, 2008)
StatusTerminated
ReasonThe bender is heating prior to activation of the power button, requiring disassembly to stop heating.
Root causeProcess control
ActionAll affected Synthes CMF Sales Consultants were notified of the Medical Device Recall by phone on 12/19/08 and by letter sent via return receipt email on 12/22/08. All affected hospitals were notified by letter mailed on 12/22/08 via USPS certified mail.
Quantity40 units
DistributionNationwide distribution: AL, CA, KS, LA, MD, NJ, NV, NY, OR, PA, TX, and UT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021458SYNTHES IN-SITU BENDER/CUTTER
Synthes (Usa), Jul 24, 2002