ClearanceBrief
Recall Z-1419-2009 ยท FLF Sterilizer, Ethylene-Oxide Gas

Advanced Sterilization Products

STERRAD 50 Sterilization System Product Code 10050 The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

DateJun 15, 2009 (initiated Mar 9, 2009)
StatusTerminated
ReasonASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs.
Root causeComponent design/selection
ActionFirm mailed out Urgent Device Correction letters dated March 9, 2009, advising customers/users that on April 23, 2008, Advanced Sterilization Products (ASP) initiated a device correction notification related to certain models of STERRAD Sterilization Systems that have the potential to emit oil mist due to a premature mechanical failure in the oil mist filter and/or filter assemblies. All affected STERRAD Systems have been corrected since the notification was issued. Recently, a secondary, lower-occurring cause of oil mist has been identified within the STERRAD NX System and STERRAD 50 System o
Quantity2,209 units installed in US; 2013 units OUS
DistributionWorldwide Distribution -- USA, Germany, Italy, France, UK, Switzerland, Benelex, Uruguay, Venezuela, Argentina, Chile, Columbia, Ecuador, Belgium, Puerto Rico, Hong Kong, Indonesia, Mexico, Philippines, Austria, China, Greece, Ireland, Israel, Japan, Korea, Malaysia, Singapore, Thailand, Taiwan, Tur
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023290STERRAD 50 & STERRAD 100S STERILIZERS
Advanced Sterilization Products, Apr 3, 2003
K981625STERRAD 50 STERILIZER
Advanced Sterilization Products, Jan 11, 1999