Recall Z-1483-2009 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Inc.

Zimmer Natural-Knee II System Modular Cemented Tibial Baseplate, Right, Size 00, for cemented use only, sterile, Zimmer, Warsaw, IN; Model Number: 6420-01-002. The Natural-Knee II System Cemented Modular Tibial Baseplate is a nonporous component which may be used in total knee replacement surgery. The baseplate is intended for cemented use only.

DateJun 30, 2009 (initiated Mar 18, 2009)
StatusTerminated
ReasonThe tibial spacer will not affix to the distal surface of the baseplate, as intended.
Root causeNonconforming Material/Component
ActionZimmer, Inc. issued an "Urgent-Device Recall" letter dated March 18, 2009 instructing users to cease using and return the affected product accompanied with a completed Inventory Return Certification form. The inventory Return Certification form should also be faxed to Zimmer, Inc. at 1-574-372-4265. Further questions should be addressed to Zimmer, Inc. at 1-800-613-6131.
Quantity834 of all products
DistributionWorldwide Distribution - USA including states of CA, NY, MI, OH, WI, IA, MN, NJ, IL, MO, TX, LA, NC, SC, PA, AR, MS, FL, GA, AL, OK, KS, TN, CT, NM, UT, NV, ID and CO and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023528NATURAL-KNEE II CEMENTED MODULAR TIBIAL BASEPLATE
Centerplus Orthopedics, Inc., Jan 17, 2003