Recall Z-1332-2009 ยท FFA Tube, Drainage, Suprapubic

Boston Scientific Corporation

Boston Scientific, Flexima and Percuflex Drainage Catheter Family (Pigtail Style) UPN DESCRIPTION M001271240 FLEXIMA APDL FIRM 10F/25CM M001271310 FLEXIMA APDL FIRM 6.3F/25CM M001271320 FLEXIMA APDL FIRM 8F/25CM M001271330 FLEXIMA APDL REG 6.3F/25CM M001271340 FLEXIMA APDL REG 8F/25CM M001271350 FLEXIMA APDL REG 10F/25CM M001271380 FL

DateMay 21, 2009 (initiated Mar 18, 2009)
StatusTerminated
ReasonBoston Scientific Corporation is initiating an urgent Field Correction related to its Flexima and Percuflex Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.
Root causeLabeling False and Misleading
ActionConsignees were sent a Boston Scientific " Urgent Medical Device Field Correction Notification" letter dated March 18, 2009. The letter was addressed to "The Hospital Field Action Contact". The letter described the problem, product affected and the revised catheter removal instructions. They also requested the return of the Acknowledgement Form.
Quantity586,276 US; 276,013 OUS total for all products
DistributionWorldwide Distribution -- USA including DC, Puerto Rico and Guam and countries of United Arab Emirates Austria, Bosnia and Herzegovina, Bahrain, Switzerland, Germany, Egypt, Spain, Finland, France, Great Britain, Ireland, Italy, Kuwait, Netherland, Pakistan, Portugal, Saudi Arabia, Sweden, Slovak Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944290CATHETERS, BILIARY CATHETERS, UROLOGICAL
Boston Scientific Corp, Dec 8, 1994