Recall Z-1123-2009 ยท GDC Table, Operating-Room, Electrical

Berchtold Corp.

Operon Electrical Operating Room Table. The device is used as an electrical operating room table.

DateApr 27, 2009 (initiated Feb 2, 2009)
StatusTerminated
ReasonDifficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.
Root causeDevice Design
ActionAn official notification letter dated February 3, 2009 was hand-carried by a local Berchtold Corporation Technical Service representative to each account. The letter indicated that the Berchtold Corporation Service Technician will complete recommended upgrades. Direct questions concerning this recall to your BERCHTOLD Technician Service at 1-800-243-5135 and press the Number 2.
Quantity453 units
DistributionWorldwide Distribution -- US and Canada.