Recall Z-0831-2011 ยท IOR Wheelchair, Mechanical

Sunrise Medical Inc

Manual Wheelchairs with Transit System Option, Model: Quickie LXI (EILXI), Product is manufactured and distributed by Sunrise Medical, Inc. Fresno, CA Intended to provide mobility to persons restricted to a sitting position and are unable to transfer from their wheelchairs when traveling in motor vehicles, the wheelchair must serve as a vehicle seat.

DateDec 28, 2010 (initiated Feb 2, 2009)
StatusTerminated
ReasonThe stabilizer bar was inadvertently omitted during the manufacturing process.
Root causeProcess change control
ActionAn Urgent: Medical Device Correction letter, dated February 2009, with certified return receipt was sent to customers in the United States; Canadian consignees were notified via e-mail. The letter identified the affected product and the reason for the recall. The letters asked customers to contact their Quickie dealer to make arrangements to have their device wheelchair corrected at no charge and to also verify their wheelchair serial number on the form. Questions should be directed towards their dealer or Sunrise Medical at 800-456-8168.
Quantity23
DistributionWorldwide Distribution -- United States and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973673QUICKIE ZIPPIE SERIES, QUICKIE BREEZY SERIES, QUICK FOLDING SERIES, QUICKIE RIGID SERIES
Sunrise Medical, Inc., Nov 20, 1997