Recall Z-1122-2009 ยท FSY Light, Surgical, Ceiling Mounted

Trumpf Medical Systems, Inc.

ILED Surgical Light, Model numbers: ILED3, ILED5 and ILED5K

DateApr 1, 2009 (initiated Feb 27, 2009)
StatusTerminated
ReasonCracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area.
Root causeOther
ActionConsignees were notified via an Urgent Safety Information letter starting 02/27/2009. The Recall Notice asks customers to visually inspect the ILED covers for cracks and in particular to focus on the corners. If cracking is visible they are asked to remedy by taping with a clear tape over the cracks and report it to TRUMPF Medical Systems, Inc. immediately. TRUMPF will be replacing all covers with a newly designed and tested cover. Original priority in the United States will be set to start with the oldest units affected. If TRUMPF receives notification that cracking does exist then that faci
Quantity4480 units (540 in U.S.)
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061317ILED
Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg, Jun 22, 2006