Recall Z-1540-2009 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Abbott Point of Care Inc.

i-STAT EG7+ Cartridge; pH, PCO, PO, Na , K, iCa, Hct. Test include: Electrode Measurement Blood-Gases (Pco2, PO2) and Blood Ph, Electrode, Ion Specific, Potassium, Electrode, Ion Specific, Calcium, Electrode, ion Specific, Sodium and Hematocrit. Catalog number 220300. Abbott Point of Care inc., Abbott Park, IL 60064 USA Intended use is to quantify measurement of sodium, potassium, ionized calc

DateJun 18, 2009 (initiated Feb 23, 2009)
StatusTerminated
ReasonEquipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals.
Root causePackaging process control
ActionUrgent Recall Notice letters pertaining to the EG7+ Cartridges were sent on February 23, 2009 by first class mail to all customers. The firm found one other affected lot of G3+ cartridges and expanded the recall to include one other lot. Letters for the expansion are dated May 2009. Customers were asked to use the attached form to record all of the box numbers if they have any cartridges from the affected lot. Cartridges from the box numbers listed will be replaced an any remaining unused cartridges from those specific boxes are to be returned as instructed on the form. To receive replace
Quantity1048 boxes of 25 cartridges each.
DistributionWorldwide Distribution -- USA, Austria, China, India, Saudi Arabia, South Africa, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K894914MODEL 100 PORTABLE CLINICAL ANALYZER AND TEST PACK
I-Stat Corp., Oct 16, 1989
K912387I-STAT MODEL 100 PORT CLIN ANALYZER & TEST PACK
I-Stat Corp., Jul 1, 1991
K934642I-STAT IONIZED CALCIUM TEST
I-Stat Corp., Dec 17, 1993
K940918I-STAT 200 PORTABLE CLINICAL ANALYZER
I-Stat Corp., May 10, 1994