Recall Z-1209-2009 ยท HRX Arthroscope

Medtronic Spine LLC, formerly Kyphon Inc

KyphX Xpander Inflatable Bone Tamp, size 15/3, Catalog Number K09A, Manufactured by Kyphon Sarl, Switzerland. For use in spinal surgery.

DateMay 11, 2009 (initiated Feb 4, 2009)
StatusTerminated
ReasonProduct not sterile: 10 units were released for distribution without undergoing sterilization process.
Root causeProcess change control
ActionSales representatives were contacted via e-mail on February 4, 2009, and instructed to contact hospitals and inform them of the issue and sequester affected product. They were further instructed to immediately travel to the hospitals and obtain affected product for return to Medtronic LLC. A Customer Communication Notice URGENT NOTICE: DEVICE REMOVAL dated February 9, 2009, was sent to the two consignees.
Quantity10 units affected.
DistributionProduct was distributed to two hospitals in OH and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041454KYPHX XPANDER INFLATABLE BONE TAMPS MODEL#KO8A,KO9A,K13A
Kyphon, Inc., Jul 9, 2004