Recall Z-1092-2009 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

GE Medical Systems, LLC

GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector) Indicated for use in diagnostic and interventional angiographic procedures of human anatomy. Intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional

DateMar 12, 2009 (initiated Mar 6, 2008)
StatusTerminated
ReasonGE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to recover.
Root causeSoftware design
ActionOn site visits were performed from 3/6/08 - 12/19/08 to update software on all 5 units.
Quantity4
DistributionWorldwide Distribution -- USA including states of MD, TX, and VA and country of China.