Recall Z-0468-2012 ยท MLR Sterilizer, Chemical

Advanced Sterilization Products

STERRAD NX Sterilizer Hydrogen Peroxide Gas Plasma Sterilizer Product Code 10133. The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

DateJan 11, 2012 (initiated Dec 10, 2008)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) has discovered a defect in the barcode labeling of the STERRAD NX System Cassettes 10133, Lot #08I032. This is the only lot number affected by this action. The barcodes for Lot #08I032 were printed improperly and are not capable of being read by the STERRAD unit.
Root causeError in labeling
ActionAdvance Sterilization Products sent an "URGENT: PRODUCT RECALL" letter dated December 10, 2008 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. The letter instructs customers to examine inventory and return all sealed cassettes, using an enclosed prepaid UPS return label to: Stericycle, 2670 Executive Drive Suite A, Indianapolis, IN 46241. Additionally, a Business Reply Card and packing slip were included with the letter for customers to complete and return. Questions or further assistance was directed to ASP Customer Care Ce
Quantity776 cases
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042116STERRAD NX STERILIZER
Advanced Sterilization Products, Apr 22, 2005