Recall Z-1206-2009 ยท MKJ Automated External Defibrillators (Non-Wearable)

ZOLL Medical Corporation, World Wide Headquarters

Zoll AED Plus Defibrillator

DateApr 1, 2009 (initiated Feb 12, 2009)
StatusTerminated
ReasonDevice fails to discharge the defibrillation energy.
Root causeProcess design
ActionOn 3/31/09, the firm revised their recall strategy and required all users to a perform a "Mandatory Upgrade" software The software upgrade allows all potentially affected devices to monitor battery charging performance through periodic self-testing. If defective batteries are detected at any time prior to the recommended maximum of five years, users are prompted by the device to install fresh batteries. Frim issued Press Release on April 2, 2009. In the first notification Zoll notified customers by letter on 2/12/09 via certified mail. Customers instructed to replace device batteries every
Quantity183,535 units
DistributionWorldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033474ZOLL AED PLUS
Zoll Medical Corp, May 21, 2004