Recall Z-1094-2009 ยท KPS System, Tomography, Computed, Emission
GE Medical Systems, LLC
GE Healthcare Infinia Hawkeye(R), Model numbers: H2508PY H3000WN H3000WR H3000WW H3000WY H3000WZ H3000YM H3000YS Intended use: To perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Onc
| Date | Apr 10, 2009 (initiated Sep 8, 2008) |
| Status | Terminated |
| Reason | There is a possibility for misregistration when brain scans are performed with Infinia with the Hawkeye option. This misregistration could present the potential for misdiagnosis when the patient headrest sags under the weight of a head. |
| Root cause | Software design |
| Action | Consignees were sent from September 2008 - Feb 2009 a GE "Product Safety Notification" letter dated 4/2008. The letter was addressed to Hospital Administrators, and Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information. It instructed consignees to continue to use the Infinia Hawkeye system normally for all types of Hybrid scans and stated that as a general practice tomograms should be reviewed carefully using the Xeleris Volumetrix processing protocol to ensure pro |
| Quantity | 153 US, 264 OUS |
| Distribution | Worldwide Distribution -- USA including states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, DC, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK ,OR, PA, PR, SD, TN, TX, UT, VT, VA, WA, WV, and WI and countries of AUSTRALIA, BELGIUM , BRASIL, CANADA, CHINA, CO |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022960 | GE QUASAR NUCLEAR MEDICINE SYSTEM Ge Medical Systems F.I. Haifa, Sep 19, 2002 |