Recall Z-1003-2009 ยท KPS System, Tomography, Computed, Emission

GE Medical Systems, LLC

GE Healthcare Infinia Hawkeye 4, Model Numbers: 5178728, ASM500060, H3000WC and H3000WD. Product is used to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology a

DateApr 21, 2009 (initiated Jun 2, 2008)
StatusTerminated
ReasonGE Healthcare has recently became aware of excessive pressure applied by the collimator pressure sensitive cover of Infinia (Infinia, Infinia Hawkeye, Infinia Hawkeye 4) that may impact patient safety. It was reported at a customer site that the pressure used to activate the Pressure Sensitive Device (PSD) should be lowered and that the body contouring software requires correction to prevent po
Root causeDevice Design
ActionConsignees were sent from June 2008 to February 2009 a GE "Product Safety Notification" letter dated April 2008. The letter was addressed to Hospital Administrators and Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information. A GE Healthcare service representative will contact each consignee to schedule the installation of the Pressure Sensitive Device (PSD). Please contact your local sales or service representative for further information regarding this recall. Fu
Quantity122 US, 166 OUS
DistributionWorldwide Distribution US (and Puerto Rico) including states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV and WI and countries of Australia, Austria, Belgium, Bra
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022960GE QUASAR NUCLEAR MEDICINE SYSTEM
Ge Medical Systems F.I. Haifa, Sep 19, 2002