Recall Z-1119-2009 ยท KPS System, Tomography, Computed, Emission

GE Medical Systems, LLC

GE Healthcare, Infinia, Model Numbers: ASM000886 ASM001391 ASM001333 ASM001465 ASM001473 ASM001620 ASM500016 H3000WM H3000WT H3000WY and SYS000041. The device is used to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging feature

DateApr 10, 2009 (initiated Jun 2, 2008)
StatusTerminated
ReasonThe metal pin in the adjustable head holder, which is used to acquire Brain SPECT data on Infinia systems (Infinia, Infinia Hawkeye, and Infinia Hawkeye 4), creates an artifact on the patient's cerebellum, which may lead to the potential for misdiagnosis.
Root causeOther
ActionConsignees were sent a GE "Product Safety Notification" letter dated April 2008. The letter was addressed to Hospital Administrators, and Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information. Contact your local sales or service representative for further information regarding this notification by calling 1-800-437-1171 (within the US or Canada) and 1-262-896-2890 (outside the US).
Quantity365 US and 408 OUS
DistributionWorldwide Distribution -- USA (DC and Puerto Rico) including states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and countries of Australia, Austria, Belg
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022960GE QUASAR NUCLEAR MEDICINE SYSTEM
Ge Medical Systems F.I. Haifa, Sep 19, 2002