Recall Z-1247-2009 ยท IYB Tube Mount, X-Ray, Diagnostic

Del Medical Systems Group

Ceiling Tube Mount Ceiling Tube Mount (CTM)/Digital Ceiling Tube Mount (DCTM) with 5 inch or 10 inch Column Extensions; a radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Del Medical Systems Group, 11550 West King Street, Franklin Park, IL 60131-1330. The device is used for X-Ray tube positioning.

DateJun 30, 2009 (initiated Jan 12, 2009)
StatusTerminated
ReasonThere is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator.
Root causeComponent design/selection
ActionDel Medical sent a "Field Correction Recall" letter with an acknowledgement form to the direct account dealers/assemblers on January 12, 2009 - February 4, 2009, informing them that required maintenance inspection of a retaining ring inside the column assembly may not be performed as required. If the retaining ring is not regularly inspected and replaced when worn or loose, there is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator. The accounts were requested to locate the affected systems and im
Quantity213 devices
DistributionWorldwide Distribution -- US, Canada, Scotland and Pakistan.