Recall Z-0949-2009 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord, CA Medical charged-particle radiation therapy system.

DateJan 30, 2009 (initiated Dec 20, 2008)
StatusTerminated
ReasonThe range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.
Root causeDevice Design
ActionOn 12/20/08, the firm issued letters via certified mail to direct consignees, informing them of the affected product and providing instructions on the recall. Contact Siemens Medical Solutions USA, Inc. at 1-925-246-8407 for additional information.
Quantity43 units
DistributionNationwide and Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072485ARTISTE MV SA
Siemens Medical Solutions USA, Inc., Dec 27, 2007