Recall Z-1198-2010 ยท HIF Insufflator, Laparoscopic

Stryker Endoscopy

Pnemosure Insufflators, model numbers 620-040-600 and 620-040-610, manufactured by Stryker Endoscopy, San Jose, CA.

DateMar 23, 2010 (initiated Dec 29, 2008)
StatusTerminated
ReasonUnexpected powering down/failure to boot up: A problem with a transistor in the heater/temperature sensor component in the device may cause the device to power off during surgery or to fail to boot up at all. The risk includes delay and interruption of ongoing surgical procedures.
Root causeComponent design/selection
ActionRecall initiated 12/29/2009. Sales representatives were notified via phone and e-mail, while product advisory notices were sent to consignees via FedEx. Consignees are provided with a response form, and are instructed to run a software correction.
Quantity884 devices
DistributionProduct was distributed throughout the US and to Argentina, Brazil, Canada, Chile and Latin America, as well as to 16 government facilities.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K06336745L CORE INSUFFLATOR, MODEL F114
W.O.M. World of Medicine AG, May 23, 2007