Recall Z-1072-2009 ยท DYB Introducer, Catheter

Boston Scientific Corp

Soft Tip Sheath Multipurpose Introducing Sheath Kit, Model Numbers: M00456600, M00456610, M00456620, M00456630, M00456650, M00456700, M00456710 and M00456760. Manufactured by Thomas Medical Products, Inc. Product is intended for use to facilitate the percutaneous introduction of various types of cardiovascular catheters into either the right or left side of the heart. The sheath may be

DateApr 9, 2009 (initiated Dec 24, 2008)
StatusTerminated
ReasonThe marker band may detach from the sheath and a clinical failure could result in an embolism.
Root causeProcess design
ActionThe firm initiated the recall on December 24, 2009 and notified consignees via "Urgent Company Initiated Medical Device Recall - Immediate Action Required" letters which explain the reason for the recall and request the affected product be returned to the Boston Scientific Corporation Distribution Center. An amendment letter with two additional Convoy Advanced Delivery Sheath Kit models (M0045748ST0 and M0045749ST0) was sent to affected customers on January 27, 2009. Direct questions about this recall to your local Sales Representative.
Quantity31,340 of all units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013866CONVOY ADVANCED DELIVERY SHEATH KIT
Boston Scientific, Dec 14, 2001
K022067CONVOY ADVANCED DELIVERY SHEATH
Boston Scientific, Sep 11, 2002
K034061MODIFICATION TO CONVOY ADVANCED DELIVERY SHEATH
Boston Scientific Corp, Jan 30, 2004
K072719CONVOY ADVANCED DELIVERY SHEATH
Boston Scientific Corp, Mar 7, 2008