ClearanceBrief
Recall Z-0984-2009 ยท HRS Plate, Fixation, Bone

Depuy Orthopaedics, Inc.

DePuy DNP Anatomic Fracture Repair System, DNP Anatomic Right, Depuy, Warsaw, IN; REF DNPAR. The device is used in the treatment of fractures and osteotomies of the distal radius.

DateFeb 11, 2009 (initiated Dec 23, 2008)
StatusTerminated
ReasonThe DNP nail plate may not attach to the alignment guide (jig) instrument, making it difficult to insert the nail plate into position or to drill needed holes accurately.
Root causeDevice Design
ActionThe firm's sales force was instructed via "Urgent Device Correction" letter issued December 29, 2008 to inspect all of these instruments in their possession, and those instruments at their hospital customers, and to leave a copy of the letter with each hospital customer who has the instruments on site. The letter describes the issue and corrective actions for customers. Please direct questions about this recall to Depuy Orthopaedics, Inc. at 888-783-7156.
QuantityApprox. 3,250 of all products
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052248DORSAL NAIL PLATE, ANATOMICAL
Hand Innovations, LLC, Sep 14, 2005