Recall Z-2113-2009 ยท LZH Pump, Infusion, Enteral

Zevex International, Inc.

Infinity Orange Enteral Feeding Pump, REF INFORNG, Zevex Enteral Nutrition Delivery Systems, Salt Lake City, UT 84123.

DateSep 21, 2009 (initiated Dec 18, 2008)
StatusTerminated
ReasonEnteral feeding pumps may not accurately deliver fluid, potentially under-infusing or over-infusing.
Root causeSoftware design (manufacturing process)
ActionAll of Zevex' customers and distributors were notified by an Urgent Medical Device Correction letter, dated 12/18/2008. The letters explained the issue, the patient impact, and what necessary actions should be taken. Customers were instructed to contact Zevex to schedule the upgrade of their installed base of pumps at 800-970-2337 or sales@zevex.com. Distributors were to contact their customers and notify them of the letter.
Quantity318
DistributionNationwide, including 1 VA facility in CA. Foreign distribution to 1 consignee in Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031199ENTERALITE INFINITY ENTERAL FEEDING PUMP
Zevex Intl., Inc., Jul 9, 2003