ClearanceBrief
Recall Z-0970-2009 ยท LLZ System, Image Processing, Radiological

GE Healthcare Integrated IT Solutions

GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS (Picture Archiving and Communication System) Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurs

DateFeb 3, 2009 (initiated Nov 20, 2008)
StatusTerminated
ReasonUnread exams may inadvertently get marked as dictated in PACS workstation when the user signs off the report in Talkstation and opens the next exam.
Root causeSoftware design
ActionGE Healthcare Integrated IT Solutions notified their Centricity PACS RA1000 Workstation with Talk Extend software customers via letter (Urgent Medical Device Correction) dated 11/10/08 of the software anomaly resulting in a patient safety issue involving Talk Extend, which could contribute to a delay of care or misdiagnosis due to delay in communicating pertinent exam results or failure to treat due to lack of an attached dictated report by the radiologist for physician review. The letter provided the users with safety instructions to follow until the software is updated. Any questions were d
Quantity22 units
DistributionCalifornia, Florida, Illinois, Iowa, Louisiana, Oregon, Pennsylvania, Texas, Utah, Virginia and Wisconsin.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043415CENTRICITY PACS SYSTEM
Ge Medical Systems Information Technologies, Jan 21, 2005