Recall Z-0969-2009 ยท LLZ System, Image Processing, Radiological

GE Healthcare Integrated IT Solutions

GE Centricity PACS RA1000 Workstation (for diagnostic image analysis); GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The Centricity PACS (Picture Archiving and Communication System) Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nur

DateFeb 3, 2009 (initiated Jan 12, 2009)
StatusTerminated
ReasonIf a user site created a custom default display protocol (DDP) to display the current exam versus the historical exam and the user-defined DDP fails, the system displays the exams using a GE default hanging protocol. This may create a potential for misdiagnosis if the user does not review the date and time stamps in the title bars.
Root causeSoftware design
ActionGE Healthcare Integrated IT Solutions notified their Centricity PACS RA1000 Workstation software customers via letter (Urgent Medical Device Correction) dated 1/12/09 of the software anomaly resulting in a patient safety issue involving Default Display Protocols. The issue may create a potential for misdiagnosis if the user does not review the date and time stamps in the title bars. The letter provided the users with safety instructions to follow until the software is updated. Any questions were directed to the Remote Online Center, Field Engineer or the GE Customer Care Center at 800-437-1171
Quantity778 units
DistributionNationwide, Canada, Bermuda, Mexico, Ecuador, Chile, and Brazil
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043415CENTRICITY PACS SYSTEM
Ge Medical Systems Information Technologies, Jan 21, 2005