Recall Z-0945-2009 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000, and E2ZCK000. IMPAX Cardiovascular is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. The IMPAX Centric Polling Procedure provides a mechanism to ensure that patient demograp

DateFeb 3, 2009 (initiated Dec 24, 2008)
StatusTerminated
ReasonPolling procedure may fail when multiple merge/split is carried out.
Root causeSoftware design
ActionAll sites potentially affected by this product were notified of the issue directly via telephone. Each site was directed to observe a ten minute polling window as an interim mitigation until the normal mitigation by software Service Update is performed at the site by Agfa Support. An Urgent Safety Notice was sent via FED-EX on/about 12/24/2008 to all sites. The letter describes the potential issue and both interim and formal mitigation. Acknowledgment, via FAX-Back, that the information was received and understood was requested from the sites.
Quantity5 units
DistributionNationwide Distribution --- including states of FL, IL, OR, SC, and SD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022292IMPAX DIAGNOSTIC DISPLAY STATION,IMPAX CLINICAL REVIEW STATION,IMPAX FOR CARDIOLOGY CLINICAL REVIEW SATATION,IMPAX ORTHO