Recall Z-0907-2009 ยท KCY Tourniquet, Pneumatic

Virtus, Inc.

Expressaire, Disposable Tourniquet Cuff 34", 1 line (18), sterile, Manufactured for Sammons Preston, A Patterson Medical Company, Cedarburg, WI.; Cuff type A70107010. Each cuff is marked ExpressAire by Sammons Preston and A70107010. The tourniquet cuff is placed on a patient's thigh during an operation procedure of the lower leg. The cuff is not operational until it is attached to an air pump s

DateJan 23, 2009 (initiated Nov 24, 2008)
StatusTerminated
ReasonThe inflation tube may come loose at the connection to the cuff during use.
Root causeEmployee error
ActionSammons Preston was notified of the recall by letter dated 11/24/08, which instructed them to examine their inventory immediately for the specific lot numbers. The quantity on hand should be indicated on the table attached to the letter and the info should be faxed to Virtus at 812-933-0749. Customers that may have tourniquet cuff inventory of identified lot numbers should be contacted. Once recall items have been located in quarantine, Tara Collins, Quality Technician, should be contacted at 812-933-1121 for Return Material Authorization number.
Quantity3654
DistributionNationwide Distribution --- including state of Wisconsin.
FDA recordOpen on fda.gov