Recall Z-0985-2009 ยท HAW Neurological Stereotaxic Instrument

Zimmer Inc.

Orthosoft Navitrack, OS Unicondylar Knee 1.0 - Universal; software application for orthopedic surgical stereotaxic instrument.

DateMar 2, 2009 (initiated Aug 19, 2008)
StatusTerminated
ReasonIncorrect parameters: If the user switches from one surgeon profile to another surgeon profile after completing anatomical landmark acquisition tasks, the application will continue to use the initial surgeon profile parameter values in the calculation of the predicted hip-knee-ankle alignment estimates, which may have different resection or component thickness parameters than are appropriate for
Root causeSoftware design
ActionZimmer notified consignees via letter dated 8/19/08, which advised them not to change the selected surgeon profile after the acquisition of landmarks has begun, and the issued will be fixed with the next software release.
Quantity75 (34 U.S. and 41 international)
DistributionWorldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Singapore, South Africa, Switzerland and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071714NAVITRACK SYSTEM - OS UNICONDYLAR KNEE UNIVERSAL, MODEL# PRO-06003
Orthosoft, Inc., Jul 20, 2007