Recall Z-0349-2009 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

Allura XPER FD20

DateSep 24, 2009 (initiated Dec 10, 2008)
StatusTerminated
ReasonTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Root causeRadiation Control for Health and Safety Act
ActionAn Electronic Product Radiation Warning, dated October 2008, was sent to customers. The letter identified the problem and under what circumstances that it could occur; the actions that should be taken by the customer/user to prevent risks for patients or users; and the actions planned by Philips to correct the problem. Further information or support concerning the issue could be obtained by contacting Bob Pettis at 425-487-7469 or the Philips Call Center at 800-722-9377. Firm FCO tracking numbers 72200107, 72200108, 72200111, 72200112, 72200117, 72200118, 72200119, 72200122, 72200123, 72200
Quantity835 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003