Recall Z-1956-2009 ยท GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
Genico, Inc., dba Genicon
Genicon, RX Only, Sterile EO. Genicon 6869 Stapoint Court, Suite 114 Winter Park, FL U.S.A. 32792. EC REP Genicon Europe, Ltd. 44 Airedale Avenue, Chiswick London U.K. W4 2NW. Catalog Number 700-005-010, 5MM. A suction irrigation device. This device is used in conjunction with an insufflator during laparoscopic surgery to provide suction and irrigation to the surgical site. The suction irriga
| Date | Jul 2, 2009 (initiated May 8, 2008) |
| Status | Terminated |
| Reason | Genicon is recalling the Genicon Disposable Suction Irrigation 5mm 700-005-010 and 10mm 700-005-005 Tube Sets. The caps on the trumpet valves of the devices could fall off, resulting in the malfunction of the Suction Irrigation Device. |
| Root cause | Nonconforming Material/Component |
| Action | All four consignees were contacted by email and fax on May 8, 2008. Wholesalers were requested to forward copies of Genicon's recall letter dated May 8, 2009 and response form to their customers. Instructions for product return was provided. Direct questions about the recall to Genicon by calling 407-657-4851 (in United States) or 0118 930 5333 (in United Kingdom). |
| Quantity | 2,790 |
| Distribution | Worldwide Distribution -- US (MO and IA), United Kingdom, Netherlands and United Arab Republic. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.