Recall Z-1042-2009 ยท JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use

Diasorin Inc.

Diasorin LIAISON (R) Chemiluminescence Analyzer Model # 15970. Colorimeter, Photometer, Spectrophotometer for clinical use.

DateMar 30, 2009 (initiated Dec 5, 2008)
StatusTerminated
ReasonDiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer.
Root causeDevice Design
ActionDiaSorin Inc authorized Field Service Engineer has visited the sites and performed the necessary field correction. Upon completion of their work, the field service engineer generated a Field Service Report and a copy was left with the customer.
Quantity9
DistributionWorldwide Distribution -- USA including states of NC, NJ, PA, CA, WA, MO, UT, and OR and country of Canada.