Recall Z-0962-2009 ยท NRA Prosthesis, Knee, Femorotibial, Unicompartmental, Semi-Constrained, Metal/Polymer, Mobile Bearing
Biomet, Inc.
Biomet Oxford Unicompartmental Knee, Phase 3, Right Medical Tibial Tray, Standard Size B, CoCr, sterile, Biomet UK Ltd, Bridgend, UK; . Unicompartmental knee joint replacement system. Designed to allow for motion between the plastic tibial bearing and metal tibial baseplate components of a unicompartmental knee prosthesis.
| Date | Feb 3, 2009 (initiated Dec 19, 2008) |
| Status | Terminated |
| Reason | Packages labeled as "right" actually contain a "left" component. |
| Root cause | Packaging process control |
| Action | Biomet salesmen were notified by phone and instructed to retrieve the products. One customer was notified by letter dated 12/19/08 that use of the product should cease and that a salesman would pick up the product. A second customer had implanted the product and was hand delivered a letter dated 12/19/08 instructing that the patient should be monitored and symptoms treated as necessary. Contact Biomet at 574-362-3983 for assistance. |
| Quantity | 2 |
| Distribution | Massachusetts and West Virginia. |
| FDA record | Open on fda.gov |