ClearanceBrief
Recall Z-0978-2009 ยท KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Boston Scientific Corporation

Gold Probe 10FR Working Length: 300 cm, Probe Tip O.D. 3.2 mm (10Fr), Rx Only. Electrical Connector: Standard. For Single Use Only. Legal Manufacturer, Boston Scientific Corporation, One Boston Scientific Place Natick, MA 01760-1537 USA. Made in USA: 8600 N.W. 41st Street Miami, FL 33166-6202 USA. The device is indicated for use in transendoscopic electrocautery of visible bleeding and nonbleedi

DateFeb 13, 2009 (initiated Dec 24, 2008)
StatusTerminated
ReasonPackaging tray for the Gold Probe Single-Use BiPolar Electrohemostasis Catheters may have a crack. Cracks in the tray or damage to the tray of the Gold Probe device may lead to a compromised sterile barrier.
Root causePackaging process control
ActionA notification letter (dated December 24, 2008) was issued to customers. The letter identified the issue. Consignees are instructed to: 1) Discontinue use and segregate affected product, 2) Complete a Reply Verification Tracking Form and fax to the Boston Scientific Corporation (BSC) at 508-683-5578 and 3) Package and ship recalled product as instructed in the letter. If affected product is found, customers are instructed to call BSC Customer Service at 1-866-868-4004 between 8:30 a.m. through 6:00 p.m. (Eastern Standard Time). Please direct questions about the recall to your local Sales R
Quantity15,334
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.