ClearanceBrief
Recall Z-0674-2009 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump, Terumo Cardiovascular Systems, Ann Arbor, MI; Model 801041. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.

DateJan 29, 2009 (initiated Nov 20, 2007)
StatusTerminated
ReasonThe system may malfunction due to t-filter corrosion on an internal circuit board.
Root causeNonconforming Material/Component
ActionU.S. consignees were notified of the problem by letter on 11/16/07. International affiliates were informed via email on 11/21/07. Consignees were informed that Terumo would schedule appointments to install new pressure modules and reconfigure the systems beginning in Feb. 2008 or that parts would be shipped to consignees wishing to perform the work themselves. For additional information, contact Terumo CVS Customer Service at 1-800-521-2818.
Quantity1219 (total)
DistributionNationwide, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003