Recall Z-2224-2009 ยท JAK System, X-Ray, Tomography, Computed

GE Medical Systems, LLC

GE Healthcare CT/e and HiSpeed X/i, families of computed tomography Systems with Gantry. The CT/e and HiSpeed X/i families CT Scanner Systems are indicated for head and whole body X-ray Computed Tomography applications.

DateSep 18, 2009 (initiated Dec 1, 2008)
StatusTerminated
ReasonDuring internal testing, GE Healthcare has recently determined that the collimator used on our CT/e and HiSpeed X/i families of Computed Tomography systems does not comply with the specific requirements in United States Title 21 of the Code of Federal Regulations, and IEC 60601-1-3, related to leakage radiation from the diagnostic source assembly, which includes the collimator. The excess leak
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare issued an "Important - Electronic Product Radiation Warning" to affected customers. GE Healthcare field service representative will perform a field corrective action on all affected units. All potential users of the device should be made aware of the notification and the information contained. For further information, contact GE Healthcare at 1-800-437-1171 (US) or by contacting your local GE Healthcare service manager.
Quantity7005 units (787 US, 6218 OUS)
DistributionWorldwide Distribution -- UNITED STATES (AL, AZ, AR, CA, CT, DC, FL, GA, IL, IN, KS, KY, LA, MD, MA, MI, LMN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, PA, PR,SC, SD, TN, TX, VA, WA, WV and WI), ALGERIA, ANTIGUA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BENIN, BOLIVIA, BOSNIA & HERZEGOVINIA, BR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021491HISPEED DUAL
Ge Medical Systems, Inc., May 22, 2002
K944013CT PROSPEED FAMILY
GE Medical Systems, Feb 7, 1995
K980169HISPEED LX//I, HISPEED FX/I, AND HISPEED DX/I FAMILY OF SYSTEMS
General Electric Co., Apr 13, 1998
K991716HISPEED NX/I CT SCANNER SYSTEM
Ge Medical Systems, Inc., Jun 18, 1999
K993645CT/E COMPUTED TOMOGRAPHY SCANNNER SYSTEM
Ge Medical Systems, Inc., Dec 21, 1999