Recall Z-0868-2009 ยท HRX Arthroscope

Stryker Endoscopy

Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows inst

DateJan 16, 2009 (initiated Dec 1, 2008)
StatusTerminated
ReasonLeak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed.
Root causeComponent change control
ActionRecall initiated 12/01/2008, Sales representatives were notified via telephone and e-mail. Product advisories entitled ""Urgent: Device Removal" and dated December 1, 2008, were sent to all consignees. All unused product is to be returned to the firm. For further information, please contact Stryker Endoscopy by telephone at 408-754-2124.
Quantity4,194 boxes (24,750 units)
DistributionWorldwide Distribution --- including USA and countries of Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, New Zealand, Poland, Portugal, Sweden, Spain, Singapore, Switzerland, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944210ARTHROSCOPE ACCESSORIES
Stryker Endoscopy, Feb 14, 1995