Recall Z-0690-2009 ยท CAI Circuit, Breathing (W Connector, Adaptor, Y Piece)

King Systems Corp.

King Systems Universal Flex2 Breathing Circuit, Latex Free, 40 per case, King Systems Corporation, Noblesville, IN; REF # CF0110-230. This breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation.

DateFeb 3, 2009 (initiated Aug 1, 2008)
StatusTerminated
ReasonThe inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.
Root causePending
ActionConsignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.
QuantityApprox. 500,000 units of all products.
DistributionWorldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.