ClearanceBrief
Recall Z-0954-2009 ยท GNW Antisera, Cf, Influenza Virus A, B, C

Millipore Corporation

LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit.

DateMar 6, 2009 (initiated Nov 24, 2008)
StatusTerminated
ReasonCertain lots of Simulfluor Flu A / Flu B Reagent may cross-react with Respiratory Synctial Virus. The continued use of Simulfluor Flu A / Flu B Reagent Catalog #5250 may lead to a potential misdiagnosis of influenza B.
Root causePending
ActionMillipore is performing a voluntary recall by notifying customers and distributors by telephone on November 24, 2008. In addition, an Urgent: Medical Device Recall notification letter with the same date was sent via certified mail. Customers were advised to examine their stock immediately and discontinue the use of the specified lots. Customers were instructed to complete the enclosed Effective Check Survey confirming disposal of defective lots and return this form to Millipore's Regulatory Department. Customer service is to be contacted at 1-800-437-7500 for replacement product. Millipore
Quantity258 units
DistributionWorldwide Distribution -- USA including states of AL, CA, CO, CT, FL, ID, IL, KS, MA, MI, MS, NM, NY, OH, OK, OR, PA, SD, TX, UT, VA, VT, WA, WV, and WI and countries of Canada, Mexico, Italy, Belgium,Germany, France, Spain, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974302LIGHT DIAGNOSTICS SIMULFLUOR FLU A/FLU B
Light Diagnostics, Apr 8, 1998