Recall Z-0862-2009 ยท GES Blade, Scalpel

Miltex, Inc.

Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.

DateFeb 20, 2009 (initiated Nov 17, 2008)
StatusTerminated
ReasonBlade may break: Over milling of the blade resulted in tip more narrow than specified.
Root causeProcess control
ActionAn Urgent Medical Device Recall letter dated 11/17/08 with an accompanying fax back form was sent to each customer. Each distributor was instructed to review their inventory, complete the form, and return any affected product. The distributors were also instructed to conduct a sub recall notification.
Quantity200 boxes
DistributionNationwide Distribution and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.