Recall Z-0622-2009 ยท GEA Cannula, Surgical, General & Plastic Surgery

Genico, Inc., dba Genicon

Genicon 5mm Pyramidal Trocar, Sterile, Genion 6869 Stapoint Court, Suite 114 Winter Park, FL U.S.A. 32792, Rx Only. Catalog Numbers: 100-005-001 & 100-005-002. Used for incisions made for positioning the laparoscopic cannulas or ports.

DateDec 24, 2008 (initiated Sep 25, 2008)
StatusTerminated
ReasonThe stainless steel tips on the Genicon 5mm Pyramidal Trocar exhibited contamination that resembled rust and there was also pitting on the tip of the trocar.
Root causeManufacturing material removal
ActionConsignees were notified with a letter, dated 9/19/2008, by e-mail and fax on 09/25/2008. The letter stated a description of the issue, analysis and reliability data, indications of Device Malfunction, and recommendations. There was also a request for all defective devices to be returned to the distributor or Genicon for replacement. Genicon Customer service was to be contacted at 407-657-4851.
Quantity2,965 units (1340 OUS and 1625 US)
DistributionWorldwide Distribution --- including USA and countries of Africa, United Kingdom, Denmark, Colombia, Belgium, Italy, Finland, Hong Kong, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982472GENICON TROCAR
Genicon, LC, Feb 4, 1999