Recall Z-2221-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use

Abbott Laboratories, Inc

ACCELERATOR APS System Input/Output Module (IOM); List Number: 7L01-01; manufactured by Inpecco SpA, Segrate (MI), Italy.

DateSep 18, 2009 (initiated Oct 27, 2008)
StatusTerminated
ReasonWhen an ARCHITECT c16000 System is connected to an ACCELERATOR APS System, the software on the ACCELERATOR APS System may not generate an error warning when a Sample Presentation/Sample Queue error occurs, which could result in contamination of subsequent samples.
Root causeSoftware design
ActionAbbott Laboratories notified Customers of the affected device via a "Product Correction" letter dated October 27, 2008. Consignees provided with a temporary work around until a service representative upgrades the system software. For further information, contact your local Abbott Laboratories, Inc. service representative.
Quantity15 units
DistributionInternational Distribution -- Spain, Italy, the United Kingdom, Sweden, Switzerland, the Netherlands, France, Canada and Scotland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.