Recall Z-0858-2009 ยท LPB Cardiac Ablation Percutaneous Catheter
Boston Scientific Corp
RF Ablation System Foot Switch. (Can be used with Boston Scientific Maestro 3000 and EPT-1000XP Cardiac Ablation Cardiac Ablation Systems). Manufactured by Altech Corporation, Flemmington NJ and distributed by Boston Scientific Corporation, San Jose, CA, catalog numbers/UPN numbers 21840/M004 218400. Indicated for use in cardiac ablation procedures.
| Date | Jan 14, 2009 (initiated Nov 25, 2008) |
| Status | Terminated |
| Reason | Product malfunction may result in delivery of energy without depressing the foot switch that controls delivery. |
| Root cause | Process control |
| Action | Recall initiated on November 25, 2008. Urgent Company Initiated Medical Device Recall - Immediate Action Required letters, dated 11/26/2008, were sent to consignees via Federal Express and included return and reply forms for consignees to respond. The letters asked that further distribution or use of any foot switch affected by this recall be ceased immediately. Attached to the letter was a document entitled "Instructions for Determining Affected Product," which provided instructions and serial numbers for determining if customers have product affected by this recall. In addition, a docume |
| Quantity | 322 units |
| Distribution | Worldwide Distribution --- including USA and countries of France, Great Britain, Saudi Arabia, Spain, Canada, Portugal, Germany, Italy, Australia, Russia, Netherlands, Ireland, Libya. |
| FDA record | Open on fda.gov |