Recall Z-2312-2010 ยท LDD Dc-Defibrillator, Low-Energy, (Including Paddles)

Physio Control, Inc.

Pediatric Paddles attachments used with LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12 and LIFEPAK 20/20e Defibrillator/Monitors.

DateAug 28, 2010 (initiated Nov 17, 2008)
StatusTerminated
ReasonSterilization efficacy could not be confirmed for sterilization methods on External Sterilizable Paddles and Pediatric Paddle attachments.
Root causeLabeling False and Misleading
ActionOn 11/14/08, Physio Control began sending the Urgent User Information letter via certified mail to their customers. The letter notified the customers about the modification to the cleaning and sterilization instruction of the External Sterilization Defibrillation Paddles and Pediatric Paddle attachments used with LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12, and LIFEPAK 20/20e Defibrillator/Monitors The customers are informed to discontinue use of the gas sterilization methods listed as 12/88 Ethylene Oxide (EO) or 100% Ethylene Oxide (OE). They are notified that they can continue to use the steril
Quantity1689 units
DistributionDevices were distributed to medical facilities through out the US and WORLD WIDE to Canada, Germany, India, Netherlands, Australia, Hong Kong, Japan, Brazil, Panama, Russia, Korea, Malaysia, Venezuela, and United Arab Emirates(UAE) .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K902288MODIFIED LIFEPAK 9 MONITOR/DEFIBRILLATOR
Physio-Control Corp., Feb 25, 1991