Recall Z-0956-2009 ยท MCM Implant, Cochlear

Advanced Bionics LLC

HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear System and Clarion Multi-Strategy Cochlear Implant are cochlear implant systems designed to provide useful hearing to individuals with

DateFeb 3, 2009 (initiated Jul 24, 2008)
StatusTerminated
ReasonMistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Potential clinical symptoms include: -No lock -Intermittent lock -Implant locks, but cann
Root causeEmployee error
ActionAdvanced Bionics is notifying clinicians of this situation so that they can provide appropriate clinical management. The notification letter sent to clinicians in the United States was sent via FedEx so that delivery information could be confirmed. The letter advises that Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Auria and Harmony headpieces are not affected by this issue. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the
Quantity2,049 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.