Recall Z-1397-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use

GE Healthcare Integrated IT Solutions

Centricity TriWin Laboratory Information System; GE Healthcare Integrated IT Solutions, Barrington, IL 60010 The Centricity TriWin Laboratory product is a laboratory information system. It is designed to facilitate the general clinical, anatomic pathology, and cytology laboratory workflow such as order entry, results entry, instrument interfacing, results reporting and patient record retrieval.

DateMay 15, 2009 (initiated Nov 17, 2008)
StatusTerminated
ReasonA software defect in the Centricity Lab (TriWin) product may incorrectly link patient information to a wrong report. Also, when viewing results in Patient Viewer, and a coagulation result is higher or lower than the linearity limit, then all of the results following the higher or lower result will show the same result as the first flagged result.
Root causeSoftware design
ActionGE Healthcare notified their Centricity TriWin Laboratory customer via letter dated 11/17/08 of the software anomalies resulting in patient safety issues involving the patient information incorrectly being linked to a wrong report, and when viewing results in Patient Viewer, and a coagulation result is outside of a linearity limit, then all of the results following the out of linearity result will show the same result as the first flagged result. The user was informed that a software patch has been prepared to resolve the issues and will be installed on their system. Any questions were directe
Quantity26 units
DistributionWorldwide Distribution -- USA and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall