Recall Z-1017-2009 ยท MNI Orthosis, Spinal Pedicle Fixation

Warsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK

Medtronic TSRH Spinal System, 3Dx Splined Connector, Large; Medtronic Sofamor Danek, Memphis, TN; REF 8351512. The device is intended for orthosis, spinal pedicle fixation for degenerative disc disease. The product is used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients; posterior non-cervical, non-pedicle screw fixation or for anterior use fo

DateMar 12, 2009 (initiated Oct 31, 2008)
StatusTerminated
ReasonThe thickness of the connector may be under specification, which could cause the implant to break during surgery or after implantation.
Root causeProcess control
ActionHospital risk managers and implanting physicians were notified separately by letters dated 1/9/09. The letters advised of the recall, reported that a firm representative would visit the facility to pick up the recalled products. Physicians were instructed to closely monitor their patients during postoperative visits, to notify any other physician who may be monitoring the patient in lieu of the implanting physician, and to advise their patients to contact them immediately if they have any sudden and significant symptoms (increased discomfort, sudden change in their clinical symptoms, or new
Quantity119
DistributionNationwide Distribution including states of Arizona, Connecticut, Georgia, Indiana, Minnesota, New Jersey, New York, South Carolina, Texas and Washington.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072317MODIFICATION TO:TSRH SPINAL SYSTEM
Medtronic Sofamor Danek USA, Inc., Sep 18, 2007