ClearanceBrief
Recall Z-1067-2009 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Corp.

Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular D Module, Roche Diagnostics Corp., Indpls., IN; GMMI No. 03284522001. Is a fully automatic, computerized chemistry analyzer intended for in vitro quantitative and qualitative determinations of a wide range analytes. The instrument is capable of performing potentiometric and photometric assays.

DateApr 10, 2009 (initiated Nov 12, 2008)
StatusTerminated
ReasonIf a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
Root causeSoftware change control
ActionConsignees were notified via an Urgent Medical Device Correction letter dated 11/12/08 and instructed to only execute "module unmasking" when the unit is in "standby" mode and to consult with their physicians at the facility to determine the clinical implications and whether any clinical intervention is needed for patients treated based upon results obtained with this software. Questions should be directed to Roche Diagnostics Technical Support at 1-800-428-2336.
Quantity22 systems in all
DistributionNationwide Distribution -- including states of California, Florida, Georgia, Hawaii, Indiana, Maryland, New York, North Carolina, Pennsylvania, and Virginia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953239BOEHRINGER MANNHEIM/HITACHI 917 ANALYZER
Boehringer Mannheim Corp., Sep 25, 1995