Recall Z-0608-2009 ยท JAK System, X-Ray, Tomography, Computed

Hitachi Medical Systems America Inc

Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician and is intended for general purpose CT applications.

DateJan 8, 2009 (initiated Nov 5, 2008)
StatusTerminated
ReasonSoftware error: The device has a software error that can result in a condition where the displayed image orientation does not match the display orientation markers (right-left reversal).
Root causeSoftware design
ActionOn 11/5/2008 Hitachi Medical Systems sent recall notification letters entitled "URGENT - DEVICE RECALL CORRECTION" to its customers informing them of the problem with the software. They stated that a Hitachi service representative will contact each customer to schedule a software upgrade. For further information, please contact Hitachi Medical Systems America Inc. by telephone at 330-425-1313.
Quantity42
DistributionNationwide Distribution --- including states of AZ, CA, DE, FL, ID, IL, IN, KS, KY, LA, MA, MO, NJ, NY, OH, OK, PA, PR, SC, TX, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040902HITACHI PRESTO
Hitachi Medical Systems America, Inc., Apr 21, 2004