Recall Z-0605-2009 ยท FSA Lift, Patient, Non-Ac-Powered

Liko North America Inc

Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Uno 100 EM/EE: S/N 7090001 through 7096200 Uno 101: S/N 10001 through 11000 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and

DateJan 27, 2009 (initiated Nov 11, 2008)
StatusTerminated
ReasonActuator of the lift has the potential to separate when the upper arm is fully extended.
Root causeProcess control
ActionHill Rom notified customers with a Field Safety Notice, dated December 5, 2008 to notify customers of the issue, potential risks, and to inspect units per their instructions and remove units from service if the unit fails inspection. Hill-Rom will provide a permanent fix, using an "OuterTube" to be placed on all units.
Quantity24070 total units
DistributionWorldwide Distribution --- including USA and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.