Recall Z-0503-2009 ยท FSA Lift, Patient, Non-Ac-Powered

Arjo, Inc.

Maxi Move, Model NDA0200-20 Batteries Pack; BHM Medical Inc., 2001 Tanguay, Magog, QC, Canada J1X 5Y5 The equipment is intended to be used in hospitals, nursing homes, or other health care facilities by trained caregivers for the transfer of patients.

DateDec 16, 2008 (initiated Nov 5, 2008)
StatusTerminated
ReasonThe battery pack NDA0200-20, lot 070826D, sold with the Maxi Move patient lift or as replacement parts, has a defective connector that could lead to the inability of the battery pack to recharge and/or short circuit, resulting in smoke emission.
Root causeComponent design/selection
ActionArjo sent the Urgent Device Recall Customer Notification dated 11/5/08 and Field Safety Notice (FSN) dated 10/31/08 to all affected Maxi Move customers on 11/5/08. The accounts were informed of the problem with the connector of the battery pack NDA0200-20, lot 070826D, sold with the Maxi Move patient lift (model numbers starting with KMC)or as replacement parts. They instructed the accounts to discontinue use of the battery pack if it does not recharge or has an intermittent electrical contact. The accounts were requested to complete and return to Arjo Inc., the enclosed customer response for
Quantity273 battery packs
DistributionNationwide Distribution