Recall Z-1006-2009 ยท MKJ Automated External Defibrillators (Non-Wearable)

Welch Allyn Protocol, Inc

Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")

DateMar 9, 2009 (initiated Dec 15, 2008)
StatusTerminated
ReasonReliability issues - potential to shock a non shockable rhythm or not shocking a shockable rhythm;
Root causeComponent design/selection
ActionWelch Allyn sent a Customer Recall Notice dated February 26, 2009, to all customers (customer list supplied by distributors). Welch Allyn will upgrade all units with software version 2.02 and lower to version 2.06.02 by Proc. AED10 Prod Service Min HW/SW Update Rev. E 820-0178-00.
Quantity10, 735
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021168JUMP START, MODEL 970300
Medical Research Laboratories, Inc., Jun 26, 2002