Recall Z-0855-2009 ยท MUJ System, Planning, Radiation Therapy Treatment

Varian Medical Systems Oncology Systems

Eclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms of the eye.

DateJan 15, 2009 (initiated Oct 2, 2008)
StatusTerminated
ReasonWhen using the multiple static segment option, the device may alter dose delivery.
Root causeSoftware design
ActionAn Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps on October 2, 2008. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. The firm recommends that customers DO NOT USE the Multiple Static Segment (MSS) option in Varian Leaf Motion Calculator version 8.2.22. The firm also recommends setting the default delivery method for Varian Leaf Motion Calculator to Sliding Window. When using Eclipse 8.5, always follow the Safety Ins
Quantity64 units
DistributionWorldwide Distribution --- including USA and countries of Switzerland and the Netherlands.